FLOMAT FX
Electromagnetic
insertion sensor
Series
Instructions Manual
R-MI-FlomatFX Rev.: 1 English version
instrumentation
for fluids
Przeglądanie stron 0
1 2 3 4 5 6 ... 11 12

Podsumowanie treści

Strona 1 - Instructions Manual

FLOMAT FX Electromagnetic insertion sensor Series Instructions Manual R-MI-FlomatFX Rev.: 1 English version instrumentation for fluids

Strona 2

7 POSSIBLE PROBLEMS WHEN COMMISSIONING 7.1 No flow rate indication Check that the cables for remote electronic units have been correctly connected.

Strona 3

8 MAINTENANCE It is recommended to clean the electrodes in installations where appreciable incrustations or sedimentations can occur. Dirty electrode

Strona 4

WARRANTY Tecfluid S.A. GUARANTEES ALL ITS PRODUCTS FOR A PERIOD OF 24 MONTHS, after consignment, against all defects in materials and workmanship. Thi

Strona 5

2 TABLE OF CONTENTS 1 WORKING PRINCIPLE …………………... 3 2 RECEPTION ...

Strona 6

3 1 WORKING PRINCIPLE The Flomat electromagnetic insertion flowmeters are based on Faraday’s induction law. When an electrically conductive liquid flo

Strona 7 - 135...145

Downstream from the sensor, the minimum recommended distance from a disturbing element is 5 x DN. 3.2 Straight pipe sections The point where the Flom

Strona 8

3.4 Valves Control valves or stop cocks should always be installed downstream from the sensor to assure that the pipe is always full of liquid. 3.3 Mi

Strona 9

6 3.6 Vibrations Vibrations of the pipes should be avoided by anchoring the pipe before and after the sensor.

Strona 10

7 For DN80 and greater, there are three lengths for each of the two types of fittings. The process of putting the insert in should be done with some

Strona 11

8 Example: Suppose a pipe of 300 mm inner diameter (DN300) and 5.5 mm thick. In the table we can see that the distance that the insertion fitting must

Strona 12

9 Drill a 48.5 mm diameter hole in the pipe to insert the fitting and weld the fitting to the pipe. The axis of the insertion fitting should be perfec