Tecfluid MX4 For FLOMAT Series Instrukcja Użytkownika Strona 9

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 8
9
4.3 Valves
Control valves or shut-off valves should always be installed downstream from the sensor
to assure that the pipe is always full of liquid.
Downstream, the minimum recommended distance to a disturbing element is 5 x DN.
4.2.1 Mixtures
If liquids of different conductivities are mixed it is necessary to install the sensor a
minimum of 50 x DN from the point of mixture in order to obtain a uniform conductivity of
the liquid and stabilize the readings.
If this distance is shorter, readings may be unstable.
4
INSTALLATION
Przeglądanie stron 8
1 2 3 4 5 6 7 8 9 10 11 12 13 14 ... 55 56

Komentarze do niniejszej Instrukcji

Brak uwag